Model Number IIPUCA4013
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994)
Date of Event 06/12/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch documentation is checked and no anomalies are detected.Any manufacturing fault is ruled out.After examination of the implant, it can be verified that the implant does not show signs of misuse or signs of malfunction that could have contributed to the event.
 
Event or Problem Description
The clinician indicates that he placed the implant on (b)(6) 2023 and 1-month later it was noted that the implant had not osseointegrated.Patient with bone quality iii.In the event the patient experienced: infection and pain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
Implant, endosseous, root-form
Manufacturer (Section D)
B.T.I. BIOTECHNOLOGY INSTITUTE S.L.
leonardo da vinci street,14
minano menor, alava 01510
SP  01510
Manufacturer (Section G)
B.T.I. BIOTECHNOLOGY INSTITUTE S.L.
leonardo da vinci street,14
minano menor, alava 01510
SP   01510
Manufacturer Contact
jose rivero
leonardo da vinci street,14
minano menor, alava 01510
SP   01510
MDR Report Key21025993
Report Number3004417597-2024-00167
Device Sequence Number15531374
Product Code DZE
UDI-Device Identifier08435389812227
UDI-Public08435389812227
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K151391
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Dentist
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberIIPUCA4013
Device Lot NumberBB01496.A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/20/2023
Initial Report FDA Received Date12/27/2024