• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC HEMOPRO2 ADAPTOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC HEMOPRO2 ADAPTOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VH-4020
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2024
Type of Reportable Event Malfunction
Event or Problem Description
The hospital reported that during an endoscopic vessel harvesting procedure, hemopro2 adaptor malfunctioned.They determined the cause to be the vh-4020 and once they switched the adapter cable, it worked without any issue.There was no delay to the case other than determining the cause and replacing the vh-4020.
 
Additional Manufacturer Narrative
Trackwise id# (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Additional Manufacturer Narrative
Tw # (b)(4).The device was returned to the factory for evaluation on 12/04/2024.An investigation was conducted on 01/22/2025.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.There were no visual defects observed on the intact adaptor.An electrical evaluation was conducted.An electrical evaluation was conducted.The adapter was connected to a reference power supply vh-3010 and reference extension cable with no visual or physical difficulties.A pre-cautery test was performed per the instruction for use with a reference hemopro 2 and reference power supply vh-3010 at level 3.0.The device passed the pre-cautery test; it did produce steam and heat during 5-second activations and shut off when the toggle was released.A polyfuse electrical test was performed with the same reference power supply, extension cable, hemopro 2 and returned adapter cable; the reference device shut off after the period of sustained activation and reactivated after 10-second cooling period with no incident each time.Based upon the evaluation results, the reported failure mode ¿electrical /electronic property problem¿ was not confirmed.The reported device is an oem device, therefore, a lot history review was not applicable.A lot/serial number was not provided and the specific product lot/serial number cannot be identified from a ship history, therefore it is impossible to obtain a certificate of conformance.
 
Event or Problem Description
The hospital reported that during an endoscopic vessel harvesting procedure, hemopro2 adaptor malfunctioned.They determined the cause to be the vh-4020 and once they switched the adapter cable, it worked without any issue.There was no delay to the case other than determining the cause and replacing the vh-4020.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOPRO2 ADAPTOR
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key21025999
Report Number2242352-2024-0001382
Device Sequence Number18391135
Product Code GEI
UDI-Device Identifier00607567700918
UDI-Public(01)00607567700918
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVH-4020
Device Catalogue NumberVH-4020
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2024
Supplement Date Received by Manufacturer02/05/2025
Initial Report FDA Received Date12/27/2024
Supplement Report FDA Received Date02/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient Age76 YR
Patient SexMale
Patient Weight93 KG
-
-