1038671-2023-01154 d10: concomitant devices (b)(6)- 02-022-35-5009 - truliant tib imp ps insert sz 5 9mm, (b)(6-,02-012-60-1425 - logic stem ext 14mm x 25mm, (b)(6)- 02-022-45-5040 - truliant tib fit tray cem sz 5f / 4t, (b)(6)- 200-07-35 - advanced patella 35mm 3 peg implant, (b)(6)- 204-70-00 - tibial stem ext.Screw.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Failure of the cement used to secure the femoral component to the femoral bone, may have lead to loosening at the cement-implant interface; however, this cannot be confirmed.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|