• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL BIOINTRAFIX TBIAL SH SM 30MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL BIOINTRAFIX TBIAL SH SM 30MM; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 254629
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unspecified surgical procedure, the biointrafix tbial sh sm 30mm device was broken.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the product has not returned to j&j medtech orthopaedics, however a photo was provided for review.¿ the photo investigation revealed that biointrafix tbial sh sm 30mm had the sheath broken in several fragments.The overall complaint was confirmed as the observed condition of the biointrafix tbial sh sm 30mm would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to axial misalignment during the sheath placement with the sheath inserter tool and/or an incorrect inserter tool size.As per the instructions for use, place the appropriate sized sheath on the corresponding sheath inserter.Refer to table 1, implant sizing guideline.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: d9, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D4 h4: the device expiration date, complete udi and date of manufacture have been added.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: investigation summary: the product was returned to j&j medtech orthopaedics for evaluation.The j&j medtech orthopaedics team conducted a visual inspection of the returned device.Visual inspection found that biointrafix tbial sh sm 30mm had the sheath broken in several fragments.Not all broken fragments were returned for evaluation.The overall complaint was confirmed as the observed condition of the biointrafix tbial sh sm 30mm would contribute to the complained device issue.¿ based on the investigation findings, the potential cause is traced to axial misalignment during the sheath placement with the sheath inserter tool and/or an incorrect inserter tool size.As per the instructions for use, place the appropriate sized sheath on the corresponding sheath inserter.Refer to table 1, implant sizing guideline.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: there were no non-conformances related to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOINTRAFIX TBIAL SH SM 30MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key21028174
Report Number1221934-2024-04647
Device Sequence Number9156226
Product Code HWC
UDI-Device Identifier10886705013258
UDI-Public(01)10886705013258(10)212L926(17)260930
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K032167
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Catalogue Number254629
Device Lot Number212L926
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/18/2024
Supplement Date Received by Manufacturer01/20/2025
01/23/2025
01/24/2025
Initial Report FDA Received Date12/27/2024
Supplement Report FDA Received Date01/22/2025
01/24/2025
01/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-