Trackwise # (b)(4).Updated section: b4, b5, d9, g3, g6, h2, h3, h6, h11.The device was returned to the factory for evaluation on 12/13/2024.An investigation was conducted on 02/17/2025.Signs of clinical use and no evidence of blood was observed.There were no visual defects observed on the intact heater wire or intact silicone insulation on both the cold and hot jaws.An electrical evaluation was conducted.A pre-cautery test was performed per the direction for use (dfu) with a reference cable, adapter, and reference power supply vh-3010 at level 3.0.The device passed the pre-cautery test; it produced visible steam and heat during ten (10) 3-second activations and shut off when the toggle was released.No excessive smoke and/or steam were observed during the testing.To evaluate the safety shut down system, a polyfuse activation test was performed 5 times over 10 minutes.The device shut off after the period of sustained activation and reactivated after 10-second cooling period with no incident each time.An activation and transection capability test was performed over four (4) repetitions using "max life test method (b)(4).The device successfully transected tissue four (4) times.A temperature and resistance test was conducted to evaluate the device function per hemopro 2 final test (b)(4).The resistance value was measured at 0.68 ohms which is within specification.The device passed the temperature measurements test.Based on the returned condition of the device as well as the evaluation results, the reported failure "electrical /electronic property problem" was not confirmed.The lot # 3000397777 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported event.
|
|
The hospital reported during an endoscopic vessel harvesting procedure, the vasoview 6 pro device not working, tried two different cords and still didn't work; device would not turn on when activated.A replacement device was used to complete the procedure.There was no procedural delay.No patient harm/effects reported.
|
|
The hospital reported during an endoscopic vessel harvesting procedure, the vasoview hemopro2 (vh-4000) device not working, tried two different cords and still didn't work; device would not turn on when activated.A replacement device was used to complete the procedure.There was no procedural delay.No patient harm/effects reported.
|