U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 4 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 4 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-35-4011
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Cyst(s) (1800); Discomfort (2330); Swelling/ Edema (4577); Joint effusion (4849)
Date of Event 12/05/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
1038671-2023-00778.D10: concomitant devices: (b)(6), 02-012-35-4011 - logic tibia ps mod insrt sz 4 11mm; (b)(6), 02-012-45-4030 - lgc tibial fit tray cem sz 4f / 3t; (b)(6), 200-02-32 - three peg patella 32mm.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported via a legal notification that a 67 yo male patient, initial left knee implanted, underwent a revision total knee arthroplasty procedure approximately 11 years post the initial procedure.The patient was having swelling and discomfort with activity.Preoperative diagnosis indicated a failed total knee arthroplasty with effusion, with concern for poly wear.Risks, complications, and alternatives were discussed, the patient agreed to the planned procedure.Incision was made.The patella was evaluated and found to have minimal if any wear.The edge of the patella was completely pristine with no evidence of loosening or debridement.There was no evidence of loosening or rocking of the proximal tibia.Attention turned to the femur.Loose femoral component with cystic formation in the medial compartment noted.Revision total knee arthroplasty, femur with a new 13 mm insert conducted.The patient was awakened and taken to recovery in good condition.No device returns anticipated due to this being a legal case.No further information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA PS MOD INSRT SZ 4 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key21028533
Report Number1038671-2024-05029
Device Sequence Number14508917
Product Code JWH
UDI-Device Identifier10885862001894
UDI-Public10885862001894
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Expiration Date06/09/2018
Device Catalogue Number02-012-35-4011
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2023
Initial Report FDA Received Date12/27/2024
Date Device Manufactured06/09/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age67 YR