| Catalog Number |
3341305 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
11/26/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation procedure with a mentor memoryshape breast implant 420cc gel breast prosthesis that possibly ruptured post implantation.Probable intracapsular rupture of the patient¿s left breast prosthesis was diagnosed via imaging.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.
|
| |
|
Additional Manufacturer Narrative
|
|
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: n/a.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 19-feb-2025, mentor received additional information about the event.The patient was scheduled to undergo explantation on (b)(6) 2025.Reason for device explant and/or reoperation: breast prosthesis rupture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 06-mar-2025, mentor received additional information about the event.The patient had both breast prostheses explanted and they were replaced with mentor memorygel boost breast implant 480cc gel breast prostheses.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 27-mar-2025, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 24-apr-2025, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient experienced a rupture in the breast implant.The product was returned to mentor for evaluation.Mentor conducted a visual inspection and microscopic examination of the returned device.Visual analysis of the returned device revealed that the breast implant was found to be ruptured and it was received in two parts.A microscopic examination was performed, and the cause of the tear could not be identified.Mentor also performed a manufacturing record evaluation related to the reported complaint for the finished device lot number.No manufacturing non-conformances were identified as part of this evaluation.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause the implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.The contralateral device was also received (lot number 7356377).The patient did not report any issue with this device.Therefore, no further analysis of the contralateral device is required.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|