Additional information was provided in d.3.,d.9.,h.3.,h.6 and h.11.Only the lens case and the lens carton were returned.The lens case was fully assembled.The lens case opened and closed securely.No damage was observed to the lens case or lens carton.The response indicated no that a company cartridge and a company handpiece were used.Because loading forceps were indicated, this would indicate that the cartridge and handpiece used were non-qualified products.The viscoelastic was also non-qualified.The root cause for the reported complaint may be related to a failure to follow the instructions for use (ifu).Non qualified products were used.The ifu instructs: company foldable intra ocular lens (iols) are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and or damage.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturers recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).
|