|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: d4 h4: the device serial/lot number, complete udi and date of manufacture are unknown at this time.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: investigation summary: the complaint device was discarded by the customer, therefore not returned to j&j medtech orthopaedics and unavailable for a physical evaluation.Given that no lot/serial number was provided, a manufacturing record evaluation (mre) review cannot be performed.If the lot/serial number becomes available, the mre review will be performed.Based on the current available information, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|