ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA STEMMED CEMENTED TIBIAL COMPONENT LEFT SIZE E; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cellulitis (1768); Unspecified Infection (1930); Skin Discoloration (2074); Swelling/ Edema (4577)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10 - concomitant devices - persona cemented all-poly patella 32mm catalog #: 42540200032 lot #: 66241175, persona posterior stabilized femoral component nitrided left size 8 catalog #: 42570606401 lot #: 65481982, persona posterior stabilized articular surface left 10mm catalog #: 42512400710 lot #: 66562667.The complainant has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient experienced cellulitis of the distal incision that extended to the foot within 3 months following left knee arthroplasty.The patient was treated with a twice-daily antibiotic for three (3) weeks, however, continues to experience swelling, discoloration and delayed wound healing.The incision is nearly healed at four (4) months following the procedure, however, the patient stated there was an indication of deep infection.Attempts have been made, however, no additional information is available.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: b4, g3, g6, h2, h3, h6, h11.The following sections were corrected: h4.The reported event of cellulitis could not be confirmed as medical records were not provided.Cellulitis is a rapid-spreading bacterial infection of the dermis and subcutaneous tissues, often caused by normal skin flora or opportunistic residential bacteria, such as streptococcus pathogens.Cellulitis appears as localized redness, swelling, and rash that is hot and painful to touch.The bacteria enter the skin via any cut, abrasion, or break in the skin, thus placing postoperative joint patients at increased risk for the development of cellulitis.Other comorbidities that increase the risk for post operative development of cellulitis include smoking, diabetes, poor nutrition, obesity, poor hygiene, or circulatory problems.Cellulitis may resolve on its own with conservative treatment but most often requires additional medical intervention, such as oral or intravenous antibiotics, to eradicate the infection.The device history records were not reviewed as the event was determined to be unrelated to the implanted device.Sterile certifications were reviewed and found to be conforming with no applicable deviations.The device was verified to have gone through acceptable sterilization processes following iso/aami/astm & eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors that may include hospital/surgical environment, provider-related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issue affecting implant safety or effectiveness, the implanted products are not identified as the source or contributing to the reported infection.The root cause was determined to be traced to non-device related factors.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA STEMMED CEMENTED TIBIAL COMPONENT LEFT SIZE E
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21034994
Report Number3007963827-2024-00541
Device Sequence Number4120664
Product Code JWH
UDI-Device Identifier00889024471054
UDI-Public(01)00889024471054(17)340527(10)66655744
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K122765
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42532007101
Device Lot Number66655744
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2024
Supplement Date Received by Manufacturer03/20/2025
Initial Report FDA Received Date12/30/2024
Supplement Report FDA Received Date03/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-