| Catalog Number |
402510 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250); Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/30/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3.Reason device not evaluated by mfg: other; device was not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a loosening or fracture at the hub between the needle and syringe, causing leakage and an inability to inject the vaccine.The operator of the device was a health professional.There was no patient involvement, and no patient harm/adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3/h6: no product was returned.The reported issue cannot be confirmed as no product was returned for investigation.If the product is returned, this complaint will be reopened for further investigation.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
|
| |
|