| Model Number |
WA02944A |
| Medical Device Problem Codes |
Break (1069); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been submitted by the importer under this mdr report number 2429304-2024-01057.The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, the ureteroscope had a bent with broken light fibers.This issue prolonged the procedure for more than 30 minutes while the patient was under sedation.The procedure was then completed with the same device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Three attempts were performed to obtain additional information, but no response was received from the customer.This supplemental report is being submitted to provide correction and the results of the legal manufacturer's final investigation.Updated fields: b5, d9, h3 h4 and h6.Corrections to b1, b2, and h1.This event was initially conservatively reported as a serious injury; however, there is no evidence the patient suffered any serious injury or adverse effects from the prolonged procedure.Olympus made multiple attempts to receive more information from the customer without success.Thus correcting b1, b2, and h1.B1 should not be marked as an adverse event but product problem only, and b2 should not be marked at all.H1 should also not be marked as serious injury but malfunction instead.The device was returned to olympus for inspection and the reported failure was not confirmed.A definitive root cause could not be identified.Based on the results of the investigation, the investigation findings of the reported failure, do not lead to a clear conclusion about the cause of the reported event.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information was received from the customer.No additional information was received from the customer.As no additional information obtained, it was determined that there was no harm to the patient as initially reported.
|
| |
|