• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH URETEROSCOPE, 7,5 FR. X 430 MM, 7°, 2 CHANNELS, STRAIGHT OCULAR ANGLE; WA02944A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH URETEROSCOPE, 7,5 FR. X 430 MM, 7°, 2 CHANNELS, STRAIGHT OCULAR ANGLE; WA02944A Back to Search Results
Model Number WA02944A
Medical Device Problem Codes Break (1069); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been submitted by the importer under this mdr report number 2429304-2024-01057.The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported that during an unknown procedure, the ureteroscope had a bent with broken light fibers.This issue prolonged the procedure for more than 30 minutes while the patient was under sedation.The procedure was then completed with the same device.There were no reports of patient harm.
 
Additional Manufacturer Narrative
Three attempts were performed to obtain additional information, but no response was received from the customer.This supplemental report is being submitted to provide correction and the results of the legal manufacturer's final investigation.Updated fields: b5, d9, h3 h4 and h6.Corrections to b1, b2, and h1.This event was initially conservatively reported as a serious injury; however, there is no evidence the patient suffered any serious injury or adverse effects from the prolonged procedure.Olympus made multiple attempts to receive more information from the customer without success.Thus correcting b1, b2, and h1.B1 should not be marked as an adverse event but product problem only, and b2 should not be marked at all.H1 should also not be marked as serious injury but malfunction instead.The device was returned to olympus for inspection and the reported failure was not confirmed.A definitive root cause could not be identified.Based on the results of the investigation, the investigation findings of the reported failure, do not lead to a clear conclusion about the cause of the reported event.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
No additional information was received from the customer.No additional information was received from the customer.As no additional information obtained, it was determined that there was no harm to the patient as initially reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETEROSCOPE, 7,5 FR. X 430 MM, 7°, 2 CHANNELS, STRAIGHT OCULAR ANGLE
Common Device Name
WA02944A
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key21037462
Report Number9610773-2024-34374
Device Sequence Number7631742
Product Code FGB
UDI-Device Identifier04042761029544
UDI-Public04042761029544
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K951855
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA02944A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/02/2024
Supplement Date Received by Manufacturer08/07/2025
Initial Report FDA Received Date12/30/2024
Supplement Report FDA Received Date08/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/29/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-