• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. GST BLACK RELOAD 45MM 6 ROW; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. GST BLACK RELOAD 45MM 6 ROW; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number GST45T
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 12/30/24.D4: batch #unk.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: did the device lock-out (no staples deployed and no cut line started)? or, did the device start to deploy staples and cut but could not be completed (partially fire)? or, did the device fully fire and stop during the automatic knife return? was there any difficulty opening the device? if yes, how was the device removed from the patient? if yes, did the jaws eventually open? attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device gst45t with lot number 349c63; and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during a right colectomy procedure, in the course of a using the psee60a, an error occurred in the process of firing.When surgeon pulled back the safety lock and fired the trigger, the blade was suddenly stopped in the middle of cartridge.As an emergency measure, the surgeon backed off the blade and detach it from the tissue.There were no adverse consequences to the patient reported nor surgical delay reported.Additional information requested.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 2/10/2025.Additional information was requested, and the following was obtained: 1.Did the device lock-out (no staples deployed and no cut line started)? no.2.Or, did the device start to deploy staples and cut but could not be completed (partially fire)? partially fired.3.Was there any difficulty opening the device? the jaw was opened manually.4.If yes, did the jaws eventually open? yes.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 07/11/2025.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 07/29/2025.D4: batch #254c63.Corrected data: d4 (udi, expiration date), h4.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one gst45t reload was received partially fired 1/10 and with an opened package.The returned reload was reset and loaded into a test device.The returned reload was tested for functionality with a test device in the straight position and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples met the staple form release criteria.It is possible that while loading the reload, it was pushed farther back than the reload alignment slot resulting in the knife pushing the one-piece sled forward and locking the reload.As part of ethicon endo surgery's quality process, all devices are manufactured, inspected, and released to approved specifications.Device history review: a manufacturing record evaluation was performed for the finished device batch number 254c63, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GST BLACK RELOAD 45MM 6 ROW
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 00969
3035526892
MDR Report Key21038221
Report Number3005075853-2024-09886
Device Sequence Number7573996
Product Code GDW
UDI-Device Identifier10705036014911
UDI-Public(01)10705036014911(10)349C63(17)251231
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K163454
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Catalogue NumberGST45T
Device Lot Number349C63
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2024
Supplement Date Received by Manufacturer02/03/2025
07/09/2025
07/27/2025
Initial Report FDA Received Date12/30/2024
Supplement Report FDA Received Date02/10/2025
07/11/2025
07/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/20/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-