| Model Number |
VTICMO 13.2 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/05/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.H6 - investigation type 4110: lens work order search - no similar complaint event(s) within associated lots were found.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.Lens exchanged with a different diopter lens.Problem resolved.Cause of the event was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation - lens was returned in microcentrifuge vial dry.Visual inspection found optic and haptic deformed.Dimensional inspection found the lens to be within specification.Functional inspection found that the returned lens did not meet original values measured at the time of manufacturing.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H6- method code 3331: device history record (dhr) review: based on the results of the investigation, all released devices from the associated work orders(s), including the suspected device, have been manufactured within established parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|