• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC NAVLOCK; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC NAVLOCK; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9734683
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/27/2023
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that there was a defective instrument due to damage.The probable cause was use of the instrument.There was no patient involvement.
 
Additional Manufacturer Narrative
The tip of one of the two side tabs had been broken off.The tracker otherwise received known good instruments without issue.With markers attached and fully seated, the tracker displayed good geometry and divot error readings with normal tracking and verification.This regulatory report is being submitted due to retrospective review through capa 626390.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVLOCK
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key21043400
Report Number1723170-2024-04086
Device Sequence Number20981017
Product Code OLO
UDI-Device Identifier00763000076634
UDI-Public00763000076634
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K171267
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 12/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9734683
Device Catalogue Number9734683
Device Lot Number190305
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2023
Initial Report FDA Received Date12/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/05/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-