MAUDE Adverse Event Report: ESTABLISHMENT LABS S.A MOTIVA ERGONOMIX ROUND SILKSURFACE; MOTIVA IMPLANTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ESTABLISHMENT LABS S.A MOTIVA ERGONOMIX ROUND SILKSURFACE; MOTIVA IMPLANTS Back to Search Results
Catalog Number ERSD-245
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/26/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Physical examination and ultrasound surveillance revealed a right breast implant rupture in a patient implanted with motiva implants.The patient was advised to undergo breast implant removal surgery.No patient demographics, such as age, weight, or ethnicity, were provided by the reporter, and no adverse outcomes for the patient have been noted.Efforts to gather additional information are ongoing, and the investigation remains in progress.
 
Additional Manufacturer Narrative
As stated in the literature, ¿a number of risk factors for rupture have been identified; the most common cause is surgical instrument damage.¿ (handel et al., 2013).Additionally, rupture is a common and known reason for complaints, and it is clearly characterized in the product dfu included with the implant, as stated above.Additionally, no cases of rupture because of product failure have been reported to establishment labs ever.Per our directions for use, each patient must receive the establishment labs s.A.¿motiva implants®: information for the patient¿ during her surgical consultation, it is the surgeons´ responsibility to ensure that the patient completely understands the information regarding the risks, benefits, and recommendations associated with silicone gel-filled breast implant surgery, as well as the complications typical of any type of surgery.This document is available in motiva® website: ifu.Motiva.Health the instructions for use in the directions for use were reviewed, to determinate if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable for the section of surgical precautions as follows: ¿breast implants can potentially remain intact for decades in the body, but all such devices will fail at some point.Breast implants rupture when the shell develops a tear or hole.Rupture can occur at any time after implantation, but it is most likely to occur; the longer the implant is in place.The following may cause implants to rupture: damage by surgical instruments, implant stress, and weakening during implantation, age and design of the implant, submuscular rather than sub-glandular location, the occurrence of post-operatory hematomas or seromas, folding or wrinkling of the implant shell, excessive force to the chest, trauma, compression during mammographic imaging, and severe capsular contracture¿.Establishment labs requested the unit to confirm the event and to perform the corresponding testing (including but not limited to: revision and characterization under the microscope of the rupture section, mechanical testing according to approved standards, etc.).It was informed that the patient/doctor has the unit and was already returned.Breast implants are not considered lifetime devices, and implant rupture is a risk associated with breast surgery, which can occur at any time.
 
Additional Manufacturer Narrative
· the reported device was returned for evaluation.There was a relationship between the reported rupture and the device.· the initial visual inspection found a rupture for the device.Microscope inspection showed trace marks in the shell consistent with those of a sharp instrument reproduced in our laboratory, this is distinct from the pattern resulting from a spontaneous tear in the shell of the implant.Elongation tests confirmed the shell complied with the international specification standards.In conclusion, it was determined that a probable cause for the rupture reported was a cut resulting from a sharp instrument used which could have weakened the shell.· establishment labs will continue to monitor for similar complaints/trends.
 
Event or Problem Description
South korea.Physical examination and ultrasound surveillance revealed a right breast implant rupture in a patient implanted with motiva implants.The patient was advised to undergo breast implant removal surgery.No patient demographics, such as age, weight, or ethnicity, were provided by the reporter, and no adverse outcomes for the patient have been noted.Efforts to gather additional information were made, and the investigation was completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTIVA ERGONOMIX ROUND SILKSURFACE
Common Device Name
MOTIVA IMPLANTS
Manufacturer (Section D)
ESTABLISHMENT LABS S.A
coyol free zone, building b25
alajuela CA 20113
CS  20113
Manufacturer (Section G)
MOTIVA USA LLC
coyol free zone, building b25
alajuela
Manufacturer Contact
susana hidalgo
coyol free zone, building b25
alajuela, CA 20113
CS   20113
8003554164
MDR Report Key21044506
Report Number3012883202-2024-00257
Device Sequence Number3493504
Product Code FTR
UDI-Device Identifier07445161010537
UDI-Public(01)07445161010537(11)190225(17)240225
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/25/2024
Device Catalogue NumberERSD-245
Device Lot Number19020308
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/30/2024
Supplement Report FDA Received Date04/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/25/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Required Intervention;
Patient SexUnknown
-
-