|
EXACTECH, INC. LOGIC FEMORAL PS CEM LEFT SZ 2; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-010-01-0220 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Osteolysis (2377); Fibrosis (3167); Implant Pain (4561); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
11/18/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
1038671-2023-00576, d10: logic tibial insert (02-012-44-2001), 3 peg patella (200-0229), logic tibial tray cemented (02-012-41-2020).Multiple mdr reports were filed for this event, please see associated reports: 1038671-2023-00576.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported via legal documentation that approximately 47 months after a left total knee replacement procedure, the patient underwent a revision procedure to address loosening, fibrosis, implant pain, osteolysis, swelling and edema.Initial surgery and revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
|
| |
|
Search Alerts/Recalls
|
|
|