It was reported that this was an incident where the foley catheter cracked.It was possible that the patient may had torn it off, but it was not broken and was not left inside the body.Per follow up information received via ibc on 18dec2024, that the crack was on the inflation funnel side of the bifurcated part.
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that this was an incident where the foley catheter cracked.It was possible that the patient may had torn it off, but it was not broken and was not left inside the body.Per follow up information received via ibc on 18dec2024, that the crack was on the inflation funnel side of the bifurcated part.
|
|
The reported event was confirmed cause unknown.Visual inspection: received sample in opened package.Sample was evaluated and found tear at bifurcation area with water leak at the tear area.This does not meet specification which states "missing, damaged, leaking, torn, broken, incomplete or incorrect components are not permitted, the clamp must be closed".No actions can be taken at this time since a root cause was not identified.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: " warnings 1.Method for use (1) do not inflate the balloon in the urethra.The urethra may be injured.(2) do not pull the catheter hard.The bladder/urethra may be injured.2.Applicable patients (1) patients with delirium who might pull out catheter.When patient tugs at catheter unconsciously, the bladder and urethra may be damaged.Contraindications 1.Method for use: (1) do not reuse.(2) do not resterilize.(3) this device contains 10 percentage povidone-iodine.For patients with past history of allergic hypersensitivity to povidone-iodine or iodine, consider using alternative disinfectants.(4) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).They may damage the device and may burst balloon.(5) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.Catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.2.Applicable patients (1) do not use in patients who are or have been allergic to natural rubber latex".Correction: d, e, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|