Catalog Number 121730052
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 12/12/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2024, the patient underwent a tha for left hip joint oa.The procedure proceeded with the regular technique.However, acetabular reaming was shallow, and sufficient fixation of the cup was not achieved.There was a crack in the femur for the stem side, so that sufficient reduction could not be done with a cable.Subsequently, the stem size was changed.The surgery was completed successfully without any surgical delay.After surgery, the patient complained of pain.The surgeon guessed that it might be due to cup loosening.On (b)(6) 2024, a revision surgery was performed.The surgeon removed the cup easily by putting in a chisel a few times.The acetabular roof was reamed again, and the cup was replaced.The stem side was not loosening, so that the stem was left as it was.The inner head was replaced with a delta ts.The surgery was completed successfully.The x-ray showed no problems.No further information is available.This pc is related to (b)(4) which reports that sufficient fixation could not be achieved during the primary surgery.This pc reports that the cup loosening occurred after the primary surgery.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MULTIHOLE W/GRIPTION 52MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21048325
Report Number1818910-2024-26220
Device Sequence Number20953305
Product Code KWA
UDI-Device Identifier10603295009986
UDI-Public(01)10603295009986(17)331130(10)M5203T
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K093646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number121730052
Device Lot NumberM5203T
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/11/2024
Supplement Date Received by Manufacturer01/02/2025
Initial Report FDA Received Date12/31/2024
Supplement Report FDA Received Date01/03/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP; CORAIL AMT SZ9 135 LOW COL; DELTA CER HEAD 12/14 36MM +1.5; PINN CAN BONE SCREW 6.5MMX15MM; PINN CAN BONE SCREW 6.5MMX20MM; PINN MAR ETO NEUT 36IDX52OD
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-