It was reported that during a posterior spinal fusion procedure from l3 to l5, while the surgeon was final tightening the lock screw at the right l4 level, the tip of the final lock screw driver fractured off within the lock screw.The tip was left in-situ, embedded in the lock screw.No further patient impact was reported.No additional information was provided.
|
|
No device was returned for evaluation; however, photographs of the device were provided for review and were able to confirm the reported event as the driver tip could be seen to be fractured flush with the point of insertion into the driver shaft.No operative notes and/or radiograph images were provided for review of usage/technique.Information regarding the torque handle used and/or the patient's bone quality was not provided.Manufacturing review discovered the device was manufactured and released to the field on sep 2019.The root cause of the fractured tip appears to be related to off angled excessive force and excessive use fatigue from 5 years of field use and repeatedly exposed to high torque situations as well as repeated sterile processing where damage can be accumulated over time.No additional investigation required.Labeling review: ".Warnings, cautions and precautions: the implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient." ".The instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices.The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury." ".Care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." ".The physical characteristics required for many instruments do not permit them to be manufactured from implantable materials.If any broken fragments of instruments remain in the body of a patient, they could cause allergic reactions or infections.If an instrument breaks in surgery and fragments go into the patient, these pieces should be removed prior to closure and should not be implanted.".
|