| Model Number |
6068093 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information received from healthcare provider via manufacturer representative regarding a patient with clinical id (b)(6), study id (b)(6) and incident type ae subevent number: 7.Primary diagnosis: instability description: slightly subsided at l4-l5 onset date: 2024-09-04 outcome status: recovering/ resolving interventions: action subtype: ae result in hospitalization, action result: no action subtype: any other action(s) taken, action result: no action subtype: did the ae result in any treatment, action result: yes action subtype: home exercise, action result: yes site seriousness assessment: there is no congenital anomaly, death, disability, hospitalization, life threatening and medical intervention, but the severity of adverse event is moderate.Site related assessment: it is not related to capstone peek spinal system, posterior supplemental fixation system, tlif grafting and study procedure.Sponsor assessment (oc muo): sponsor assessed event as unlikely related to procedure, tlif grafting material, posterior supplemental fixation system and possible related to intervertebral body fusion device.Diagnostics: action type: diagnostic action subtype: x-ray1 action result: intact hardware posteriorly l4-s1.Severe disc space narrowing with endplate sclerosis l1-2.3mm retrolisthesis l1 on l2.2mm retrolisthesis l3 on l4.6mm anterolistesis l4 on l5.Action date: (b)(6)2024 action type: diagnostic action subtype: were any diagnostic test performed action result: yes event: it was reported that patient had syncopal episode at home after surgery and prior to 6 week follow-up.X-ray images were obtained according to protocol noting that at l4-l5 subsided as compared to immediate post-op images prior to discharge.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 has been updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information was received via manufacturer representative that no revision surgery has been scheduled or planned for the removal of subsided cage at l4-l5.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information was received that description: anterolistesis l4 on l5.
|
| |
|