Investigation: visual inspection: during the investigation, scratches on the outer housing of the valve.Permeability test: a permeability test has shown that all components are permeable.Computer controlled test: to investigate the claim of over- drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the m.Blue is operating within the accepted tolerance in the vertical position, but not within the specified tolerances in the horizontal position.A slower outflow could be determined.Adjustment test: the m.Blue valve was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection: after dismantling of the valve, deposits were found.Results based on our investigation results, we can determine a slower outflow of the m.Blue in the horizontal position.The determined deposits can be named as the cause for the functional deviation.Organic material in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.
|