| Catalog Number |
314-13-34 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Inflammation (1932); Pain (1994); Loss of Range of Motion (2032); Discomfort (2330); Osteolysis (2377); Joint Laxity (4526); Muscle/Tendon Damage (4532); Swelling/ Edema (4577); Aseptic loosening (4883)
|
| Date of Event |
03/08/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: ((b)(6)) 310-01-50 - equinoxe, humeral head short, 50mm (beta).((b)(6)) 300-10-45 - equinoxe replicator plate 4.5mm o/s.((b)(6)) 315-35-00 - glnd kwire.((b)(6)) 300-01-12 equinoxe humeral stem primary press fit 12mm.((b)(6)) 315-35-00 - glnd kwire.((b)(6)) 300-20-02 equinox square torque define screw drive kit.The product associated with the reported event is within the scope of recall z-1415-2024; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/ technique.A definitive root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported via legal documentation that approximately 60 months after a left total shoulder replacement procedure, the patient underwent a revision procedure to address prosthesis wear, pain, discomfort, inflammation, impaired range of motion; component part loosening; soft tissue damage and bone loss.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.H6: health effect - clinical code, medical device problem code.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2025-02249.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, loosening, loss of range of motion, or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not returned to exactech for evaluation and no images, radiographs, or clinical information were provided.B3: corrected.D6a: implant date unknown.D6b: explant date corrected.H6: corrected investigation clinical codes.
|
| |
|