Additional Manufacturer Narrative
|
|
Multiple mdr reports were filed for this event, please see associated reports: 1038671-2023-00077.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, femoral loosening, migration/subsidence and fracture or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported via legal documentation that approximately 136 months after a left total knee replacement procedure, the patient underwent a revision procedure to address prosthesis fracture, ambulation difficulties, balance problems, failure of implant, osteolysis, joint laxity, synovitis and implant pain.Initial surgery and revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
|
| |
|