• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND, INC EPIQ ELITE; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND, INC EPIQ ELITE; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC Back to Search Results
Model Number EPIQ ELITE ULTRASOUND SYSTEM
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/14/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the swivel mechanism of the epiq elite ultrasound system¿s control panel arm will not lock into position while transporting the unit within the user facility.No patient or user was harmed because of the issue.A philips service engineer removed the control panel arm covers and found the bushing around the lower swivel lock latch was out of place.The philips service engineer cleaned the grease and residue from the hole where the bushing seats and applied loctite 480 to the bushing and seated it firmly in place to resolve the issue.A thorough engineering investigation has been performed to identify the root cause of the reported issue.The engineering team determined the failure was a hardware issue in the articulating arm latching mechanism preventing the locking solenoid from fully engaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIQ ELITE
Common Device Name
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
roxanne ramirez
22100 bothell everett highway
bothell, WA 98021
MDR Report Key21057342
Report Number3019216-2025-000001
Device Sequence Number15470900
Product Code IYN
UDI-Device Identifier00884838102514
UDI-Public(01)00884838102514
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K200304
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 01/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEPIQ ELITE ULTRASOUND SYSTEM
Device Catalogue Number795234
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2024
Initial Report FDA Received Date01/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-