DEPUY IRELAND - 9616671 XTNDGLENO D42MM +0MM SHRTPST; SHOULDER PROSTHESIS, REVERSE CONFIGURATION
|
Back to Search Results |
|
| Catalog Number |
130761142 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that intraoperatively it was not possible to engage the glenosphere with the metaglene.Two glenosheres did not engaged with the metaglene.No adverse patient consequences reported.No surgical delay was reported.To complete the surgery a glenosphere 42+4 was implanted together with the initial metaglene.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary- no device associated with this report was received for examination.A manufacturing record evaluation were performed for the finished device 130761142, lot number d24090676, it was manufactured on 03-sep-2024.21pcs parts were manufactured per specification and all raw materials met specification, and no non-conformances / manufacturing irregularities were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot- a manufacturing record evaluation were performed for the finished device 130761142, lot number d24090676, it was manufactured on 03-sep-2024.21pcs parts were manufactured per specification and all raw materials met specification, and no non-conformances / manufacturing irregularities were identified.
|
| |
|
Search Alerts/Recalls
|
|
|