|
MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507200BC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
11/14/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation was performed for the finished device 5692690 number, and no non-conformances were identified.Manufacturing date: jul/21/2006, expiration date: jul/20/2011.A manufacturing record evaluation was performed for the finished device 5624630 number, and no non-conformances were identified.Manufacturing date: aug/15/2005, expiration date: aug/14/2010.Reason for device explant and/or reoperation: anisomastia, material rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent primary breast augmentation with two 200cc mentor memorygel breast implants.Post-operatively, the patient needed a breast lift on an unspecified breast.As a result, the patient underwent breast lift surgery on (b)(6) 2024.During the surgery, the implant was identified to be ruptured.Subsequently, the patient's implants were replaced bilaterally with the following devices: (left) 200cc mentor memorygel breast implant catalog: 3502001bc lot: 9993927 sn: (b)(6) and (right) 200cc mentor memorygel breast implant catalog: 3502001bc lot: 9993927 sn: (b)(6).Since it was not specified, which breast had the complications, both the left and right implants information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
On february 17, 2025, the mentor failure analysis lab received the device for evaluation.Mentor became aware that the suspect medical device was the left breast side.On february 21, 2025, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor conducted visual inspection of the returned device.Visual analysis of the returned device revealed that the gel mod-rnd 200cc breast implant was found to be ruptured and received in three parts.In addition, an area of shell abrasion was observed on the edges of the tear.These findings are consistent with normal wear of the device.Shell abrasion suggests in-vivo folding or creasing of the device, which may be caused by continuous and sustained stresses to the device, such as an excessively small breast pocket or the folding or wrinkling of the shell within the breast pocket.In some cases, breast implants may also experience normal wear over time.The contralateral device was also received (lot number 5692690).The patient did not report any issue with this device.Therefore, no further analysis of the contralateral device is required.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.
|
| |
|
Search Alerts/Recalls
|
|
|