• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL CERENOVUS ENTERPRISE; INTRACRANIAL NEUROVASCULAR STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL CERENOVUS ENTERPRISE; INTRACRANIAL NEUROVASCULAR STENT Back to Search Results
Catalog Number ENC452812
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section h3 - the device is available to be returned for evaluation and testing.However, it has not been received to date.If the device returns, a device investigation will be performed.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
As reported by the field, during an endovascular embolization, after a non j&j microcatheter (mc) was in place, an eu 4.5x28mm stent 12 mm dw tip intracranial stent (enc452812, 9020416) passed through the microcatheter and started to release.Suddenly, the microcatheter migrated, then the physician retracted the stent into the microcatheter and adjusted the microcatheter position.The physician delivered the stent to release it again, but the stent was impeded at the tip of the microcatheter and could not pass through the microcatheter.The doctor removed the stent and microcatheter from the patient and switched to new devices to complete the procedure.The procedure was prolonged about 10 minutes.Additional event information received on 20-dec-2024 indicated that they were able to torque the device.There were no evidence of physical material within the device.The microcatheter did not kink/bent.There was no excessive force used with the devices.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).The product has been returned for evaluation and testing; however, the engineering evaluation has not been completed.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Updated sections on this med watch report: b4, g3, g6, h2, h3, h6 and h11.Complaint conclusion: as reported by the field, during an endovascular embolization, after a non j&j microcatheter (mc) was in place, an eu 4.5x28mm stent 12 mm dw tip intracranial stent (enc452812, 9020416) passed through the microcatheter and started to release.Suddenly, the microcatheter migrated, then the physician retracted the stent into the microcatheter and adjusted the microcatheter position.The physician delivered the stent to release it again, but the stent was impeded at the tip of the microcatheter and could not pass through the microcatheter.The doctor removed the stent and microcatheter from the patient and switched to new devices to complete the procedure.The procedure was prolonged about 10 minutes.Additional event information received on 20-dec-2024 indicated that they were able to torque the device.There were no evidence of physical material within the device.The microcatheter did not kink/bent.There was no excessive force used with the devices.A non-sterile eu 4.5x28mm stent 12 mm dw tip was received contained in the decontamination pouch.Upon receiving the device, visual inspection was performed, and no appearance of damages was observed.All the components were received in good condition.Further inspection was performed under a microscope, the delivery wire was inspected along its entire length, and no damages were found.The stent was found still inside the introducer.The stent was observed in good condition, with no structural damage (i.E., no broken struts, no kinks).A functional test was performed.A lab sample of the prowler select plus was flushed using a lab syringe.The unit was then inserted into the prowler select plus, and it advanced smoothly without any resistance or friction as the stent passed through.The stent was successfully expelled from the distal tip of the microcatheter.As the functional test was performed without issues, the impeded condition encountered during the procedure could not be confirmed.However, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the analysis.Lake region medical did review the device history records relative to the manufacturing, inspecting and packaging of the lot 9020416.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As part of the j&j medtech quality process, all devices are manufactured, inspected, and released to approved specifications.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no capa activity is required at this time.It should be noted that multiple factors could cause product failure.The instructions for use (ifu) do contain the following recommendations: ¿ the introducer must be properly engaged with the infusion catheter hub to enable stent introduction into the infusion catheter.¿ do not apply undue force if resistance is encountered at any point during stent manipulation.Withdraw the unit and advance to a new one.¿ if resistance is felt while recapturing the stent, do not continue to recapture the device.Withdraw the infusion catheter slightly to unsheathe the stent (without exceeding the recapture limit), and then attempt to recapture the stent again.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERENOVUS ENTERPRISE
Common Device Name
INTRACRANIAL NEUROVASCULAR STENT
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS INC
325 paramount dr
raynham MA 02767
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key21060104
Report Number3008114965-2025-00002
Device Sequence Number12957466
Product Code NJE
UDI-Device Identifier10886704044031
UDI-Public(01)10886704044031(17)260428(10)9020416
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
H60001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/28/2026
Device Catalogue NumberENC452812
Device Lot Number9020416
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2024
Supplement Date Received by Manufacturer01/16/2025
01/21/2025
Initial Report FDA Received Date01/02/2025
Supplement Report FDA Received Date01/17/2025
01/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MICROCATHETER (UNSPECIFIED BRAND).
Patient SexUnknown
-
-