• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC FEMORAL PS CEM RIGHT SZ 5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC FEMORAL PS CEM RIGHT SZ 5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-010-01-0350
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Synovitis (2094); Joint Laxity (4526); Implant Pain (4561); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 03/16/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Report number: 1038671-2023-02045, 1038671-2025-00033 concomitants: (b)(6), 02-012-45-5050 - lgc tibial fit tray cem sz 5f / 5t, (b)(6), 200-02-35 - three peg patella 35mm, (b)(6), 02-012-44-5011 - logic tibia implant psc insert, sz 5, 11mm.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2023-02045, 1038671-2025-00033.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, cracking, femoral loosening, and tibial loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
As reported via legal documentation, a patient had right knee replacement surgery on (b)(6) 2012.They underwent right knee revision surgery on (b)(6) 2023, approximately 10 years 8 months after their initial procedure.Patient experienced loosening, joint laxity, implant pain, synovitis and failure of implant.On (b)(6) 2023 op report intraoperative findings include loose femoral and tibial implants - removed easily with moderate bone loss.There was extensive soft tissue debris throughout the knee joint consistent with synovitis from a polyethylene reaction to the tissue.There was tiny, small balls of light brown synovium along with a "pseudo-capsule" of dense soft tissue.The surgeon noted there was excessive wear with signs of oxidation / yellow discoloration, pitting, cracking, and delamination of the polyethylene.Distal femoral and anterior femoral bone loss was moderate.Tibial bone loss was extensive and required placement of a tibial cone.There is no other patient demographic or medical history available.There is no device return.There are no photos or other images of the device provided.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC FEMORAL PS CEM RIGHT SZ 5
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
MDR Report Key21060450
Report Number1038671-2025-00035
Device Sequence Number18176791
Product Code JWH
UDI-Device Identifier10885862001269
UDI-Public10885862001269
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/27/2017
Device Catalogue Number02-010-01-0350
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2023
Initial Report FDA Received Date01/02/2025
Date Device Manufactured03/01/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age74 YR
Patient SexMale
-
-