| Catalog Number |
400174 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B.3.The date received by manufacturer has been used for this field.H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd syringe nrfit¿ lok had mixed product / lots.The following information was provided by the initial reporter: not nrfit connection of ref no (b)(4), wrong in production line! when did the incident occur? before use.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Mixed product / lots.Thirty-nine samples were provided to our quality team for investigation.All syringes were found with nrfit scale markings on a luer-lok barrel.The condition observed is non-conforming per product specification.Potential root causes for the mixed product defect could be associated with failure to properly contain any previously conveyed product prior to the nrfit run and inadequate line clearance on the marking machinery.Situation analysis and corrective and preventative actions were initiated with multiple corrective actions identified and implemented.Additionally, a corrective action opened under internal exception was to better control the process related to conveyed raw materials.These changes will be documented in updates to the process control forms and applicable work instructions.Operators and supervisors signed off to the machinery and line clearance documentation were re-educated to the documentation via internal system.Furthermore, a device history record review was completed for provided lot number 1041220.A review showed no rejected inspections or quality issues during the production that could have contributed to the reported defect.
|
| |
|
Event or Problem Description
|
|
No additional information was provided.
|
| |
|
Search Alerts/Recalls
|