| Model Number |
VTICM5_12.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
09/22/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H3: lens was returned with liquid in a microcentrifuge vial.Visual inspection found a haptic bent lens.Dimensional and functional inspection found lens to manufactured within specifications.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A4-a6:unk.H6: work order search: no similar complaints within associated lots were found.Manufacturer narrative: each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.The lens was later removed and replaced for a same size, different model and power lens.The problem was resolved.Cause of the event is unknown.
|
| |
|
Search Alerts/Recalls
|