Model Number BI70002000
Medical Device Problem Codes Failure to Advance (2524); Application Program Freezes or Fails to Launch (4031)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(h3,h6) no parts have been received by the manufacturer for evaluation.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the gantry would not open or close when pendant button was pressed.Neither the open or close button were working.No patient was present at the time of event.
 
Additional Manufacturer Narrative
Medtronic representative went to the site to test the imaging system and replaced door cable winch assembly.Upgraded door slides.Completed system checkout.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: bi71000429, serial/lot #: unknown, product id: bi71000273, serial/lot #: unknown.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
No add info received.
 
Additional Manufacturer Narrative
Section e updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SANMINA -SCI SYSTEMS
km 15.5 no. 29, plant 06
carr. chapala-guadalajara,jal 45640
MX  45640
Manufacturer (Section G)
SANMINA -SCI SYSTEMS
km 15.5 no. 29, plant 06
carr. chapala-guadalajara,jal 45640
MX   45640
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key21064350
Report Number3006544299-2025-00005
Device Sequence Number8484164
Product Code OWB
UDI-Device Identifier00763000068066
UDI-Public00763000068066
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K200074
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBI70002000
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/26/2024
Supplement Date Received by Manufacturer01/08/2025
01/17/2025
Initial Report FDA Received Date01/03/2025
Supplement Report FDA Received Date01/15/2025
02/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/24/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-