• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. TRULIANT PS CEM FEM PS CEM LEFT SZ 3.5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRULIANT PS CEM FEM PS CEM LEFT SZ 3.5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-020-11-0235
Medical Device Problem Codes Unstable (1667); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994); Scar Tissue (2060); Synovitis (2094); Joint Laxity (4526); Swelling/ Edema (4577); Joint effusion (4849)
Date of Event 12/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via legal documentation that approximately 30 months after a left total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, loosening, instability, scar tissue, synovitis, joint effusion, joint laxity, inflammation, swelling/ edema, and pain.Revision surgery operative notes were provided.Patient left the operating room in stable condition.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
1038671-2023-00755 d10: concomitants: (b)(6), 02-020-11-0235 - truliant ps cem fem ps cem left sz 3.5.(b)(6), 02-022-45-3525 - truliant tib fit tray cem sz 3.5f / 2.5t.(b)(6), 02-029-99-1001 - fluted headless pin 3.0" square head.(b)(6), 02-029-99-1002 - threaded head pin 2.3" square head.(b)(6), 200-07-32 - advanced patella 32mm 3 peg implant.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2023-00755.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT PS CEM FEM PS CEM LEFT SZ 3.5
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key21068239
Report Number1038671-2025-00050
Device Sequence Number8409439
Product Code JWH
UDI-Device Identifier10885862314444
UDI-Public10885862314444
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K170240
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number02-020-11-0235
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2023
Initial Report FDA Received Date01/03/2025
Date Device Manufactured03/30/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age51 YR
Patient SexFemale
-
-