Catalog Number 67PFSS40
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/27/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H3.Reason device not evaluated by mfg: other; device was not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that the cuff did not inflate.The event happened at the hospital and was operated by a health professional.The issue was resolved by using a different trach.There was patient involvement, and patient injury was reported, but no additional details were given.
 
Additional Manufacturer Narrative
H6: evaluation codes updated.Device evaluation: no product was returned.Based on the review of information provided from the customer and no device was returned for review, a product evaluation cannot be performed to confirm the reported problem.If the product is returned at a later date, the complaint will be reopened and an investigation will be completed.The device history record of reported lot number 4459605 was reviewed and it was confirmed that no non-conformities were reported during production.The quality inspection forms were reviewed, and it was observed that no defects were found, and the lot was released.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TIGHT-TO-SHAFT PEDIATRIC FLEXTEND STRAIGHT NECK TRACHEOSTOMY TUBE
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V.
ave calidad no. 4, parque
tijuana
MX 
Manufacturer Contact
reed covert
6000 nathan ln n
minneapolis, MN 55442
2247062300
MDR Report Key21068407
Report Number9617604-2025-00006
Device Sequence Number8373040
Product Code JOH
UDI-Device Identifier10351688518699
UDI-Public(01)10351688518699(17)290320(10)4459605
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number67PFSS40
Device Lot Number4459605
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2024
Supplement Date Received by Manufacturer01/23/2025
Initial Report FDA Received Date01/03/2025
Supplement Report FDA Received Date02/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age11 MO
Patient SexUnknown
-
-