| Catalog Number |
123522A |
| Medical Device Problem Code |
Inaccurate Flow Rate (1249)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Pain (1994); Urinary Tract Infection (2120); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/13/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the critical care nurses reported in an online survey that the patient experienced adverse events bacteriuria or urinary tract infection, urine retention directly caused attributable to usage of the device bardia silicone elastomer coated latex 2-way, 10ml foley catheter, 22 fr.And resulted in patient injury following this medical or surgical intervention for bacteriuria or urinary tract infection cause pain, for urine retention bold.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the critical care nurses reported in an online survey that the patient experienced adverse events bacteriuria or urinary tract infection, urine retention directly caused attributable to usage of the device bardia silicone elastomer coated latex 2-way, 10ml foley catheter, 22 fr.And resulted in patient injury following this medical or surgical intervention for bacteriuria or urinary tract infection cause pain, for urine retention blood.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.It is unknown whether the device had met relevant specifications.The product was used for urological care.It was unknown whether the product had caused the reported failure.The potential root cause for this failure could be user related (example: salt accumulation)/block drainage lumen/no drainage eye).No actions can be taken at this time since a root cause was not identified.The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: warning: do not use ointments or lubricants having a petroleum base.They will damage the latex.Visually inspected the product for any imperfections or surface deterioration prior to use.-use luer tip syringe to inflate with state 10ml of sterile water or - for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.For urological use only.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe "stick" in the valve.Allow the pressure within to balloon force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|