EXACTECH, INC. 145-DEG PE 46MM HUM LINER +2.5; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Catalog Number |
320-46-03 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
06/07/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Approximately 7 month(s) and 13 day(s) post-operative of a revised right rtsa, it was reported via clinical study that the patient experienced disassociation of glenosphere with increased pain and significant poly wear.The patient was previously revised for aseptic glenoid loosening.The patient underwent a standard reverse with preserve stem revision for the disassociation with the reported removal of the humeral liner, glenosphere, and glenoid base plate components.The outcome of this event is considered resolved and the case report form indicates that this event is possibly related to the device and/or to the procedure.This was reported through clinical trial study data collection activities, no device return is anticipated, and no other information is available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
(d10) concomitant devices: 300-30-06 - equinoxe preserve stem 6mm: 6683371.320-10-05 - equinoxe reverse tray adapter plate tray +5: 6670279.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: h6, h11.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/ technique.A review of manufacturing data was unable to be performed as the lot information of the product involved in the event was not available.A review of complaint history was unable to be performed as the relevant device and event information was unknown.A definitive root cause was unable to be determined; however, may have resulted from prosthesis wear and disassembly.However, the reported prosthesis wear and disassembly could not be confirmed.Additionally, potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: b3, d1/d2a/d2b.The revision reported may have been due to wear and disassembly.However, the reported prosthesis wear and disassembly could not be confirmed.Additionally, potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|