• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. BRONCHOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-H1100
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the bronchovideoscope experienced a b30 scope communication error and an e226 scope communication error.The issue occurred during setup/inspection for use.There were no reports of patient harm.
 
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected fields: d4, d8, d9, h2, h3, h4, h6, h11.The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: e226 confirmed due to failed connector plug unit and circuit board unit.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was reported the bronchovideoscope experienced a b30 scope communication error and an e226 scope communication error.The issue occurred during setup/inspection for use.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRONCHOVIDEOSCOPE
Common Device Name
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key21069118
Report Number9610595-2025-00046
Device Sequence Number3493464
Product Code EOQ
UDI-Device Identifier04953170424229
UDI-Public04953170424229
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K222861
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBF-H1100
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/05/2024
Supplement Date Received by Manufacturer02/04/2025
Initial Report FDA Received Date01/03/2025
Supplement Report FDA Received Date03/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/08/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-