• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC TIBIA TRAPTRAY CEM SZ 5F/4T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC TIBIA TRAPTRAY CEM SZ 5F/4T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-41-5040
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Failure of Implant (1924); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Implant Pain (4561); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 06/20/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via legal documentation that approximately 121 months after a left total knee replacement procedure, the patient underwent a revision procedure to address loosening, ambulation difficulties, discomfort, failure of implant, implant pain and osteolysis.Revision surgery operative notes were provided.Patient left the operating room in stable condition.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
1038671-2023-00893 d10: 02-012-35-5013 - 1452760 - logic tibia ps mod insrt sz 5 13mm, 02-012-35-5013 - 1452760 - logic tibia ps mod insrt sz 5 13mm, 02-010-01-0250 - 2279605 - logic femoral ps cem left sz 5.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2023-00893.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, femoral loosening, and osteolysis or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of manufacturing, user, and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and product information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA TRAPTRAY CEM SZ 5F/4T
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL
Manufacturer Contact
geoff gannon
2320 nw 66th ct
MDR Report Key21069493
Report Number1038671-2025-00056
Device Sequence Number19652713
Product Code JWH
UDI-Device Identifier10885862003256
UDI-Public10885862003256
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/22/2016
Device Catalogue Number02-012-41-5040
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2023
Initial Report FDA Received Date01/03/2025
Date Device Manufactured02/22/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age82 YR
Patient SexMale
-
-