• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO PDB; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO PDB; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number OMS-PDBS2
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/09/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, on three devices, the balloon did not inflate.It did not hold air.
 
Additional Manufacturer Narrative
D10 concomitant product: oms-pdbs2, oms-pdbs2 pdb balloon x5 (lot#p3e0406); oms-pdbs2, oms-pdbs2 pdb balloon x5 (lot#p2l0219); 24055- , 24055 5.5 sht sec prt 5.5mm lck cnn/trc (lot#j4f2714gy).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the balloon would not inflate.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDB
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key21072145
Report Number2647580-2025-00048
Device Sequence Number12855271
Product Code GCJ
UDI-Device Identifier10884521067943
UDI-Public10884521067943
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K946002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOMS-PDBS2
Device Catalogue NumberOMS-PDBS2
Device Lot NumberP4A1081
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/10/2024
Supplement Date Received by Manufacturer02/24/2025
Initial Report FDA Received Date01/03/2025
Supplement Report FDA Received Date03/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-