• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 WITH PED.BURRHOLE RESERVOIR; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 WITH PED.BURRHOLE RESERVOIR; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX490T
Medical Device Problem Codes Mechanical Problem (1384); Obstruction of Flow (2423)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 12/07/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that a progav 2.0 (#fx490t) was implanted during a procedure performed on (b)(6) 2024.According to the complainant, at the patient's follow-up checkup on (b)(6) 2024, the clinical consultant had trouble reprogramming the valve, initially set to zero (0) centimeters (cm) h20.The valve was eventually adjusted to one (1) cm h20.However, on (b)(6) 2024, the surgeon reported that the patient was experiencing hydrocephalus symptoms consistent with underdrainage.Further examination revealed a kink in the peritoneal catheter.The patient underwent a revision procedure on (b)(6) 2024.The complaint product was sent to the manufacturer for investigation.No patient complications were reported as a result of the revision procedure.The adverse event is filed under aic reference (b)(4).
 
Additional Manufacturer Narrative
Investigation: visual inspection: during the investigation, no significant deformation or damage of the valve was determined.Permeability test: a permeability test has indicated that the valve has a blockage.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in the horizontal position.Because the valve is not permeable, the computer control test is not applicable.Adjustment test: the progav 2.0 was tested and is not adjustable.Braking force and brake function test: the brake functionality test has shown that the brake function is operational.However, the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection: after dismantling of the valve, organic deposits were found.Summary of the investigation results: based on our investigation results, we can determine an occlusion and a non-adjustability of the valve.The visible deposits in the valve have led to the functional impairment.Deposits caused by substances naturally present in the patient's body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic material can affect the integrity of the valve.No further actions are required as a result of the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 WITH PED.BURRHOLE RESERVOIR
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
MDR Report Key21072436
Report Number2916714-2024-00214
Device Sequence Number20925923
Product Code JXG
UDI-Device Identifier04041906137786
UDI-Public(01)04041906137786
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/30/2025
Report Date (Section F) 01/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX490T
Device Catalogue NumberFX490T
Device Lot Number2007965
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date12/27/2024
Event Location Hospital
Date Report to Manufacturer12/07/2024
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/03/2025
Supplement Report FDA Received Date01/30/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexMale
Date Report Sent to FDA01/03/2025
-
-