EXACTECH, INC. LGC TIBIAL FIT TRAY CEM SZ 2F / 2T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-45-2020 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Synovitis (2094)
|
| Date of Event |
07/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
1038671-2023-01654 d10: 02-010-03-0220 - (b)(6) - logic cr femoral cem, left sz 2.200-02-29 - (b)(6) - three peg patella 29mm.02-012-47-2013 - (b)(6) - logic cr tib insert std, sz 2, 13mm.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2023-01654.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear or due to inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported via legal documentation that approximately 50 months after a left total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear and synovitis.Revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
|
| |
|
Search Alerts/Recalls
|
|
|