Catalog Number 010001
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/11/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: internal complaint reference: case (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that, during a shoulder surgery, the probe straight tip broke.Surgery was completed with a back-up device instead.There was no surgical delay, and no further complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the reported device was received for evaluation.A visual inspection of the returned device finds the device returned with the distal tip fractured off and not returned.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include excessive force applied during use.Based on this investigation, there is no evidence to suggest a design, material or manufacturing issue.Therefore, the need for corrective action is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROBE STRAIGHT
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key21076609
Report Number1219602-2025-00019
Device Sequence Number20956768
Product Code NBH
UDI-Device Identifier03596010021533
UDI-Public3596010021533
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number010001
Device Lot Number51216471
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2024
Supplement Date Received by Manufacturer02/05/2025
Initial Report FDA Received Date01/05/2025
Supplement Report FDA Received Date02/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/28/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexMale
Patient Weight83 KG
-
-