• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC ATELLICA CH; ATELLICA CH CALCIUM_2 (CA_2)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC ATELLICA CH; ATELLICA CH CALCIUM_2 (CA_2) Back to Search Results
Catalog Number SMN 11097644
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported to siemens that they obtained discordant elevated calcium_2 (ca_2) results on multiple patient samples on an atellica ch 930 analyzer.The elevated results were reported to the physician and were questioned.The samples were not repeated, and the correct results for the patients are unknown.The customer did not provide any patient sample data.There are no known reports of patient intervention or adverse health consequences due to the elevated calcium_2 (ca_2) results.
 
Additional Manufacturer Narrative
A united states (us) customer contacted the siemens healthcare diagnostics remote services center (rsc) regarding discordant elevated calcium_2 (ca_2) results obtained on multiple patient samples on an atellica ch 930 analyzer.Siemens is evaluating the event.
 
Additional Manufacturer Narrative
Additional information (07-feb-2025): siemens healthcare diagnostics concluded the investigation of the event.Siemens reviewed the information provided by the customer and instrument data files.Quality control (qc) recovered higher when processed using specific calibration but within the customer¿s expected ranges, indicating that the calcium_2 (ca_2) assay was performing acceptably.Siemens determined that there was no issue with performance of the atellica ch 930 analyzer.A potential product issue was not identified.The cause of the event is unknown.The customer is operational.The device is performing within specifications.No further evaluation is required.Section h6 has been updated to reflect the siemens investigation.Initial mdr 2432235-2025-00010 was filed on 06-jan-2025.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATELLICA CH
Common Device Name
ATELLICA CH CALCIUM_2 (CA_2)
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
RANDOX LABORATORIES LTD.
registration # 8020890
55 diamond road
crumlin co. antrim BT29 4QY
UK   BT29 4QY
Manufacturer Contact
varsha joshi
pritech park sez, block 11
sarjapur outer ring road
karnataka 56010-3
IN   560103
870577433
MDR Report Key21078276
Report Number2432235-2025-00010
Device Sequence Number8443961
Product Code CJY
UDI-Device Identifier00630414220697
UDI-Public(01)00630414220697(10)242227(17)250401
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K083386
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2025
Device Catalogue NumberSMN 11097644
Device Lot Number242227
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2024
Supplement Date Received by Manufacturer02/07/2025
Initial Report FDA Received Date01/06/2025
Supplement Report FDA Received Date03/04/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/01/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-