| Catalog Number |
0279351400 |
| Medical Device Problem Code |
Failure to Deliver Energy (1211)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device had insufficient / no energy delivered.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: the serfas probe stopped working after approx.60 seconds.Patient was not affected.The operation could be completed with another probe.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause could be fluid ingress reaching the pcb board, which resulted in a short circuit, likely due to a leak from a broken/damaged bond of the suction tubing.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the device had insufficient / no energy delivered.
|
| |
|