| Model Number |
2120 |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: unknown; no information has been provided to date.E: phone number extension (b)(6).H3: the device was received for evaluation.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.Visual and functional tests were performed, which found a delaminated downstream occlusion (dso) seal.The dso seal and sensor assembly were replaced to fix the issue.
|
| |
|
Event or Problem Description
|
|
The results of the investigation found that the device had a delaminated downstream occlusion (dso) seal.There was unknown patient involvement.
|
| |
|