• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD-SOLIS VIP AMBULATORY INFUSION PUMP; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD-SOLIS VIP AMBULATORY INFUSION PUMP; PUMP, INFUSION Back to Search Results
Model Number 2120
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3: unknown; no information has been provided to date.E: phone number extension (b)(6).H3: the device was received for evaluation.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.Visual and functional tests were performed, which found a delaminated downstream occlusion (dso) seal.The dso seal and sensor assembly were replaced to fix the issue.
 
Event or Problem Description
The results of the investigation found that the device had a delaminated downstream occlusion (dso) seal.There was unknown patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-SOLIS VIP AMBULATORY INFUSION PUMP
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key21079700
Report Number3012307300-2025-00263
Device Sequence Number8333742
Product Code FRN
UDI-Device Identifier10610586042836
UDI-Public(01)10610586042836(11)220321
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K111275
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2120
Device Catalogue Number21-2120-0102-78
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2024
Initial Report FDA Received Date01/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/21/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-