| Model Number |
SUTR ANCH,BIO-COMP P-LCK,2.9X 12.5MM |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Additional information received on 12/18/20254: this was discovered during a labral repair procedure.The anchors were removed from the patient.The case was completed using an ar-3636 knotless fibertak.The case was delayed twenty minutes without additional anesthesia.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3, h3, h6.Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 12/9/2024, it was reported by a sales representative via email that the eyelets of three ar-2923bc biocomposite pushlock eyelets broke off while barely inserting the anchor into the bone.Everything was removed from the patient.The case was completed using 1.8 knotless fibertak devices.This was discovered during a procedure on (b)(6) 2024.Additional information requested 12/16/2024.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to improper bone prep; misaligned insertion; prying/leveraging the device during insertion.
|
| |
|
Search Alerts/Recalls
|