ARTHREX, INC. SUTR ANCH,BIO-COMP P-LCK,2.9X 12.5MM; BIO SOFT TISSUE FIXATN FASTNR Back to Search Results
Model Number SUTR ANCH,BIO-COMP P-LCK,2.9X 12.5MM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/09/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 12/9/2024, it was reported by a sales representative via email that the eyelets of three ar-2923bc biocomposite pushlock eyelets broke off while barely inserting the anchor into the bone.Everything was removed from the patient.The case was completed using 1.8 knotless fibertak devices.This was discovered during a procedure on (b)(6) 2024.Additional information requested 12/16/2024.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
Additional information received on 12/18/2024: this was discovered during a labral repair procedure.The anchors were removed from the patient.The case was completed using an ar-3636 knotless fibertak.The case was delayed twenty minutes without additional anesthesia.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to improper bone prep; misaligned insertion; prying/leveraging the device during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTR ANCH,BIO-COMP P-LCK,2.9X 12.5MM
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21080329
Report Number1220246-2025-09224
Device Sequence Number12845425
Product Code MAI
UDI-Device Identifier00888867099203
UDI-Public(01)00888867099203(11)230516(17)250430(10)15093345
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K101679
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model NumberSUTR ANCH,BIO-COMP P-LCK,2.9X 12.5MM
Device Catalogue NumberAR-2923BC
Device Lot Number15093345
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2024
Supplement Date Received by Manufacturer12/18/2024
06/16/2025
Initial Report FDA Received Date01/06/2025
Supplement Report FDA Received Date01/14/2025
07/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/16/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-