TITAN SPINE ENDOSKELETON¿ TAS INTERBODY SYSTEM; AWL Back to Search Results
Model Number 2300-1120
Medical Device Problem Codes Break (1069); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis #(b)(4) lot# y201005 after visual and optical examination and functional testing, it does not appear to be any damage, nor does it indicate any functional issues with the awl.The handle appears to be worn from repeated normal use.H11: this regulatory report is being submitted due to retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from multiple sources (manufacturer representative, user facility) regarding an instrument used for spinal therapy.It was reported that handle found to rotate around instrument during routine servicing.It was found broken in spd.There was no patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSKELETON¿ TAS INTERBODY SYSTEM
Common Device Name
AWL
Manufacturer (Section D)
TITAN SPINE
6140 w. executive dr. suite a
mequon WI 53092
Manufacturer (Section G)
TITAN SPINE
6140 w. executive dr. suite a
mequon WI 53092
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key21088584
Report Number3006340236-2025-00004
Device Sequence Number15175647
Product Code HWJ
UDI-Device Identifier00191375001959
UDI-Public00191375001959
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2300-1120
Device Catalogue Number2300-1120
Device Lot NumberY201005
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2023
Initial Report FDA Received Date01/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-