|
It was reported that the patient was being revised due to complaint of pain.The primary components were an attune size 7 ps femur, rotating platform tibial base plate, a size 7 x 10 mm insert and 32mm anatomic patella.During the case, it was discovered that the patella component was loose.The surgeon removed the patella component, cleaned off the bone re-drilled the peg holes and implanted a 32 mm anatomic patella.He also trialed a thicker tibial insert to address instability and ultimately implanted a 14 mm size 7 insert.Unknown surgical delay.Doi: unknown.Dor: on (b)(6) 2024.Affected side: unknown.
|
|
Product complaint # (b)(4).Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.H3, h6: no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|