It was reported that they received a complaint of a defective product of ribbed balloon catheter.The foley catheter did not have a hole in the end, leading the staff to have to re-catheter the patient.There were no adverse events related to the defective product of use.However, the used the product was discarded and not retained for the complaint process.Per customer via phone on (b)(6)2025, it was reported that the item was thrown away.There was no injury to the patient.
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
The reported event was inconclusive.The conditions of the sample did not allow further evaluation.Visual evaluation of the returned one-photo sample noted one opened (without original packaging), used latex foley catheter.Visual inspection of the one-photo sample noted that the foley catheter balloon was inflated.Due to the poor condition of the photo sample the reported failure was unable to be evaluated.No actions can be taken at this time since a root cause was not identified.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and states the following: sterile unless package is opened or damaged.Warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage latex and may cause the balloon to burst.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Caution: this product contains natural rubber latex which may cause allergic reactions.Single patient use only.Do not reuse.Do not resterilize.Recommended inflation capacities: 3cc balloon: use 5ml sterile water, 5cc balloon: use 10ml sterile water, 15cc balloon: use 20ml sterile water, 20cc balloon: use 25ml sterile water, 30cc balloon: use 35ml sterile water, 40cc balloon: use 45ml sterile water, 75cc balloon: use 80ml sterile water.Do not exceed recommended capacities.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Correction- d, e, h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that they received a complaint of a defective product of ribbed balloon catheter.The foley catheter did not have a hole in the end, leading the staff to have to re-catheter the patient.There were no adverse events related to the defective product of use.However, the used the product was discarded and not retained for the complaint process.Per customer via phone on 03jan2025, it was reported that the item was thrown away.There was no injury to the patient.
|